SPETEC® Cleanroom Cell
Cleanroom technology can be used in the manufacturing, packaging, or storage of dust-sensitive mechanical and electronic components and everywhere where a particle-free environment is required.
Spetec GmbH has now expanded its cleanroom offerings to a total of three different cleanroom concepts. This makes it possible to realize all requirements, from cost-effective solutions to high-quality cleanrooms according to GMP guidelines for pharmaceuticals.
The cleanroom module consists of anodized aluminum profiles and can be installed in any laboratory, production room/hall, or office without the need for structural modifications. A complete cleanroom workstation can also be installed within the cleanroom module. Thanks to the profile technology, the size can be freely chosen between 3 m² and approximately 350 m².
The module itself is framed by a odorless strip curtain or side panels made of acrylic or laminated glass or aluminum composite panels, creating a sealed unit. Additional options such as electric sliding doors, material or personnel lock systems, with mutual locking if needed, are possible.
By combining individual laminar flow modules of various sizes, the effective cleanroom area can be determined flexibly. Thus, even production lines or conveyor belts can be integrated into the cleanroom module.
For the SPETEC® cleanroom module, H14 filters are used with a cleanroom class of 100. These have an efficiency of 99.995%. This means that the filter filters out at least 99.995% of all particles at a particle size of 0.12 μm (according to MPPS). At a particle size of 0.3 μm, the efficiency is approximately 99.9995%. The laminar airflow under the flow modules prevents exchange with contaminated outside air.
Using a SPETEC® cleanroom module, the particle concentration can be reduced from approximately 15 million/m³ (size 0.12 μm, MPPS) to about 1,500 particles within the unit. This corresponds to a cleanroom class DIN ISO 5 or class 100 according to US Fed. 209E.
Additionally, ISO classes 6 to 9 (DIN ISO 14644-1) and classes A to D according to GMP can be realized. Thus, SPETEC® cleanroom modules can be custom-designed and adapted to specific requirements depending on the application.
![]()