- Translated with AI
Kalibrator for low differential pressures
With the Micro Cal 869 SETRA of system presents a portable Kalibrator,
which is able to generate lowest differential pressures with an
accuracy from 0,05 Pa to. This compact system works perfectly
independently of net or compressed air supplies and can be used thus
at any place. Principal item of this Kalibrators is a patented
pressure generator with 2 pressure chambers. This produces the desired
differential pressure in shortest time without significant
overshooting. Within the smallest offered range from +-50 Pa a
dissolution of 0,025 % is reached. There is available ranges to 5000
Pa. The operation 869 been made by an inserted PDA, the software leads
the user fast and reliably by all necessary inputs, like measuring
range and accuracy of the test specimen, number der Kalibrierpunkte or
einschwingzeiten.Over the integrated printer the user receives a comprehensive
documentation of the calibration procedure.
PCB Piezotronics GmbH
41836 Hückelhoven