- Translated with AI
Dokuschleuse
The MCRT Micro CleanRoom Technology GmbH developed an active small
article air-lock (CLAPS). It serves for the particle cleaning of small
articles, as tool and documents such as DIN A4 file or the like, which
are to be transferred already prepurified by the grey area into the
white range. The high internal change of air ensures within an
adjustable time of e.g. 30 sec. for the dismantling of the loosely
adhering particles Abreinigungsstation with closes the door the
circulating air fan for 60 sec. is activated and the door electrically
locked. At the same time a red signal lamp shines for the indication
of the enterprise. At expiration of the pre-programmed cleaning
duration the fan switches off, which concerns green light and the door
knows is opened and the cleaned property taken.Durchreichschleuse this system differs from the one-sided
Abreinigungsstation by the actual air-lock function. That means it
that the property which can be cleaned is brought likewise prepurified
into the air-lock and at expiration o.g. Abreinigungszeit by the pure
space side to be taken can. The controlling of the Durchreichschleuse
essentially takes place similar to the Abreinigungsstation. When
opening the pure side no fan start takes place, but the tueroeffnung
depends on the grey space thus the Einbringseite. A green red
announcement permits optical control also from the pure side. Options
for both variants light barrier for the monitoring of the interior of
the air-lock. The light barrier releases the enterprise of the fans
and thus the cleaning cycle only if really property is in the air-lock
present. With empty chamber the cleaning cycle does not start.Particle counter into the technique range of the document air-lock is
inserted a 2-Kanal particle sensor with integrated vaccum pump. This
supervises the particle concentration during the cleaning cycle of the
document air-lock underneath the property which can be cleaned. A
release to open the air-lock door takes place only after a defined
particle number. Optionally the particle values in 2 size channels on
separate displays can be indicated. Ionization into the air flow of
the document air-lock can be integrated an ionization system, around
possible ESD of problems to eliminate www.mcrt.de
MCRT Micro CleanRoom Technology GmbH
35452 Heuchelheim
Germany